Installation of the ProbeUsing the documents pertaining to the magnet whose field is to be measured, estimate the value of the field. Make sure that this field is in the probe's measurement range. At one end of the probe box there is a (+) to show where the resonant material is located ( Simply tape the top side of the probe (perpendicular to the z-axis, see In addition, prevent that part of the probe cable which is located in the magnetic field from moving. Installation of the Pre AmplifierIf possible, attach the pre amplifier to the magnet. Electrically isolate the pre amplifier from the ground. Connect the probe cable to one side of the pre amplifier and the cable to the gaussmeter on the other side ( Installation of the Gaussmeteron the front panel:
on the back panel:
|
||
|
|
|